FLIR A320 25º thermal imaging camera

This product is discontinued.
This is the model that replaces it:
FLIR A615, Apliter Termografia

FLIR A615 Thermal Imaging Camera

Consult price

Description

FLIR A320 thermal imaging camera
automation with integrated processor

The FLIR A320 thermal imaging camera is a cost-efficient and accurate solution for temperature monitoring systems using thermography. The camera incorporates a web service configurable processor, so it can autonomously control temperature monitoring applications with simple control logic and very fast setup.

Outstanding images and high thermal sensitivity

The FLIR A320 thermal camera is capable of detecting small temperature variations thanks to the high sensitivity of the uncooled microbolometer detector, which provides sharp radiometric images and accurate temperature measurements.

Multiple connectivity for installation

Multiple FLIR A320 thermal imaging cameras can be connected to monitor different points in your facility or process via Ethernet connection and SNTP protocol. Each thermal camera is assigned a unique IP address for detection and data exchange. Alarm messages and images can be sent via e-mail using the same protocol.

Easy to set up and operate

The FLIR A320 thermal imaging camera features an intuitive menu system that allows users to select and manipulate multiple temperature setpoints, measurement range, image color palettes, multiple alarms and more - quickly and easily. It's as simple as connecting the camera to a PC monitor and immediately viewing the thermal image in real time. Using the IR Config software and IR Monitor it is possible to configure all camera parameters.

Ref. 48201-1201, 42701-1201

Technical data

Imaging and optical data

IR resolution
320 × 240 pixels
Thermal sensitivity/NETD
< 0.05°C @ +30°C (+86°F) / 50 mK
Field of view (FOV)
25° × 18.8°
Minimum focus distance
0.4 m (1.31 ft.)
Focal length
18 mm (0.7 in.)
Spatial resolution (IFOV)
1.36 mrad
Lens identification
Automatic
F-number
1.3
Image frequency
30 Hz
Focus
Automatic or manual (built in motor)
Zoom
1-8× continuous, digital, interpolating zooming on images

Detector data

Detector type
Focal Plane Array (FPA), uncooled microbolometer
Spectral range
7.5-13 µm
Pitch detector
25 µm
Detector time constant
Typical 12 ms

Measurement

Object temperature range
-20 to +120°C (-4 to +248°F)
0 to +350°C (+32 to +662°F)
Accuracy
±2°C (±3.6°F) or ±2% of reading

Measurement analysis

Spotmeter
4
Area
4 boxes with max./min./average/position
Isotherm
1 with above/below/interval
Measurement option
Measurement Mask Filter
Schedule response: File sending (ftp), email (SMTP)
Difference temperature
Delta temperature between measurement functions or reference temperature
Reference temperature
Manually set or captured from any measurement function
Atmospheric transmission correction
Automatic, based on inputs for distance, atmospheric temperature and relative humidity
Optics transmission correction
Automatic, based on signals from internal sensors
Emissivity correction
Variable from 0.01 to 1.0
Reflected apparent temperature correction
Automatic, based on input of reflected temperature
External optics/windows correction
Automatic, based on input of optics/window transmission and temperature
Measurement corrections
Global and individual object parameters

Alarm

Alarm functions
6 automatic alarms on any selected measurement function, Digital In, Camera temperature, timer
Alarm output
Digital Out, log, store image, file sending (ftp), email (SMTP), notification

Set-up

Color palettes
Color palettes (BW, BW inv, Iron, Rain)
Set-up commands
Date/time, Temperature°C/°F

Storage of images

Storage media
Built-in memory for image storage
File formats
Standard JPEG, 16-bit measurement data included

Ethernet

Ethernet
Control, result and image
Ethernet, type
100 Mbps
Ethernet, standard
IEEE 802.3
Ethernet, connector type
RJ-45
Ethernet, cable
CAT-6 cable standard accessory
Ethernet, communication
TCP/IP socket-based FLIR proprietary
Ethernet, video streaming
MPEG-4, ISO/IEC 14496-1 MPEG-4 ASP@L5
Ethernet, image streaming
16-bit 320 × 240 pixels:
- Signal linear
- Temperature linear
- Radiometric
Ethernet, power
Power over Ethernet, PoE IEEE 802.3af class 0
Ethernet, protocols
TCP, UDP, SNTP, RTSP, RTP, HTTP, ICMP, IGMP, ftp, SMTP, SMB (CIFS), DHCP, MDNS (Bonjour), uPnP

Digital input/output

Digital input, purpose
Image tag (start/stop/general), Input ext. device (programmatically read)
Digital input
2 opto-isolated, 10-30 VDC
Digital output, purpose
As function of ALARM, Output to ext. device (programmatically set)
Digital output
2 opto-isolated, 10-30 VDC, max 100 mA
Digital I/O, isolation voltage
500 VRMS
Digital I/O, supply voltage
12/24 VDC, max 200 mA
Digital I/O, connector type
6-pole jackable screw terminal

Composite video

Video out
Composite video output, PAL and NTSC compatible
Video, standard
CVBS (ITU-R-BT.470 PAL/SMPTE 170M NTSC)
Video, connector type
Standard BNC connector

Power system

External power operation
12/24 VDC, 24 W absolute max
External power, connector type
2-pole jackable screw terminal
Voltage
Allowed range 10-30 VDC

Environmental data

Operating temperature range
-15°C to +50°C (+5°F to +122°F)
Storage temperature range
-40°C to +70°C (-40°F to +158°F)
Humidity (operating and storage)
IEC 60068-2-30/24 h 95% relative humidity +25°C to +40°C (+77°F to +104°F)
EMC
  • EN 61000-6-2:2001 (Immunity)
  • EN 61000-6-3:2001 (Emission)
  • FCC 47 CFR Part 15 Class B (Emission)
Encapsulation
IP 40 (IEC 60529)
Bump
25 g (IEC 60068-2-29)
Vibration
2 g (IEC 60068-2-6)

Physical data

Weight
0.7 kg (1.54 lb.)
Camera size (L × W × H)
170 × 70 × 70 mm (6.7 × 2.8 × 2.8 in.)
Tripod mounting
UNC ¼"-20 (on three sides)
Base mounting
2 × M4 thread mounting holes (on three sides)
Housing material
Aluminium

Box contents

  • FLIR A320 thermal imaging camera with 25° lens
  • Ethernet cable
  • Mains cable
  • Power cable, pigtail
  • Power supply
  • Printed documentation

Related products